Series 2600B System SourceMeter® Instrument Reference Manual Section 7: TSP command reference2600BS-901-01 Rev. B / May 2013 7-215Also seesmuX.measure.autorangeY (on page 7-198)smuX.measure.Y() (on page 7-213)smuX.source.autorangeY (on page 7-217)smuX.trigger.source.limitY (on page 7-247)smuX.trigger.source.linearY() (on page 7-248)smuX.trigger.source.listY() (on page 7-249)smuX.trigger.source.logY() (on page 7-250)Sweep Operation (on page 3-20)smuX.nvbufferYThis attribute contains a dedicated reading buffer.Type TSP-Link accessible Affected by Where saved Default valueAttribute (R) Yes Not applicable See Details Not applicableUsagebufferVar = smuX.nvbufferYbufferVar The dedicated reading bufferX Source-measure unit (SMU) channel (for example, smua.nvbuffer1 applies toSMU channel A)Y SMU nonvolatile buffer (1 or 2)DetailsEach SMU channel contains two dedicated reading buffers: smuX.nvbuffer1 and smuX.nvbuffer2.All routines that return measurements can also store them in either reading buffer. Overlapped measurementsare always stored in a reading buffer. Synchronous measurements return either a single-point measurement orcan be stored in a reading buffer if passed to the measurement command.The dedicated reading buffers can be saved to internal nonvolatile memory (to retain data between powercycles) using the smuX.savebuffer() function.Examplesmua.measure.overlappedv(smua.nvbuffer1) Store voltage readings from SMU channel Ainto SMU channel A dedicated readingbuffer 1.Also seeConfiguring and running sweeps (on page 3-29)Reading buffers (on page 3-6)savebuffer() (on page 7-161)smuX.makebuffer() (on page 7-197)smuX.measure.overlappedY() (on page 7-209)smuX.savebuffer() (on page 7-216)smuX.trigger.measure.action (on page 7-241)smuX.trigger.measure.set() (on page 7-241)smuX.trigger.measure.stimulus (on page 7-242)smuX.trigger.measure.Y() (on page 7-244)