Self-Test ErrorsThe following errors indicate failures that may occur during a self-test.Refer to the Service Guide for more information.601 Front panel does not respond602 RAM read/write failed603 A/D sync stuck604 A/D slope convergence failed605 Cannot calibrate rundown gain606 Rundown gain out of range607 Rundown too noisy608 Serial configuration readback failed609 DC gain x1 failed610 DC gain x10 failed611 DC gain x100 failed612 Ohms 500 nA source failed613 Ohms 5 uA source failed614 DC 1000V zero failed615 Ohms 10 uA source failed5Chapter 5 Error MessagesSelf-Test Errors179