Environmental conditionsDataParameterTemperature: +15 to +30°COperating conditionsHumidity: 30 to 70% (no dew condensation)Temperature: -10 to +40°CNon-operating condi-tions Humidity: 20 to 70% (no dew condensation)Temperature: -25 to +70°CTransportation & stor-age conditions (within96 hours) Humidity: 10 to 80% (no dew condensation)151 W/h (reader block + FLA Image Eraser)Heat radiationIt is recommended to lower the lighting level to about 20lux when moving a sample from the cassette onto thePhosphor stage after exposure. 20 lux is the brightnesslevel at which most people can read newspaper-sized textafter he/she becomes accustomed to the brightness.LightingIndoor use onlyLocation of use2,000 m above sea levelMaximum altitude foruseTransient overvoltage category IIOvervoltage categoryPollution Degree 2Applicable PollutionRatingNoise levelsDataParameter70 dB (A) or lowerNoiseIP20Degrees of protectionprovided by enclosureGetting Started with Typhoon FLA 7000 28-9607-64 AD 117F Specifications