26Displayz This indicator means “the balance is measuring calibration data”.Do not allow vibration or drafts to affect the balance while thisindicator is displayed.7-2 Automatic Self Calibration (Calibration due to changes in temperature)This function automatically calibrates the balance when the balance detects an ambient temperaturechange. If GLP output is selected in the function table, the balance outputs the calibration report orstores the data in memory. Automatic self calibration functions even if the display is turned off(standby state).CautionIf something is on the weighing pan, the balance judges that it is in use and does not performautomatic self calibration. To maintain the calibrated state, keep the weighing pan clear whilenot in use.The displays shown below are related to the automatic self calibration.Indicates that the balance detects a change in ambienttemperature and automatic self calibration will start. If the balanceis not used for a few minutes with this indicator blinking, thebalance performs automatic self calibration. The blinking durationdepends on the environment.Indicates that the balance is measuring calibration data. Do notallow vibration or drafts to affect the balance while this indicator isdisplayed. After calibration, the balance returns to indicate theprevious display.NoteThe balance can be used while the indicator blinks. But, it is recommended that to maintainthe accuracy, stop using the balance and confirm that there is nothing on the pan and allowthe balance to perform self calibration.