5614. CONNECTION TO PERIPHERAL EQUIPMENT14-1 Connection to the AD-8121B PrinterSet the following parameters to use the AD-8121B printer.Example of use AD-8121B mode settingTo print A&D standard format weight data, using the FX-i WP PRINT keyor FX-i WP auto print mode. (The time and date can be added.) MODE 1To print A&D standard format weight data, using the AD-8121B DATA keyor AD-8121B built-in timer. (The time and date can be added.)To print, using the AD-8121B chart printing function.MODE 2To print the FX-i WP statistical data. MODE 3To print GLP output. MODE 3Class Item andParameterFactorysettingsAD-8121BMODE 1AD-8121BMODE 2AD-8121BMODE 3prtData output mode 0 0,1,2,4,5 *1 3 0,1,2,4,5 *1doutData output pU5eData output pause 0 0 0 0,1 *2bp5Baud rate 2 2 2 2btprData bit, parity bit 0 0 0 0CrlfTerminator 0 0 0 05ifSerial interfacetypeData format 0 0 0 1*1 Set appropriate parameters for “ap-p (Auto print polarity)” and “ap-b (Auto print difference) “ whenauto print mode A or B (prt 1 or 2) is selected.Set the AD-8121B DIP switch No.3 to ON when unstable data is printed with “prt 4“.*2 Set 1 when multiple lines are printed in the FX-iWP statistical calculation mode.NotesRefer to "10-2 GLP Report" for print samples.Settings of AD-8121B DIP switchesMODE AD-8121B DIP switch DescriptionMODE 11 2 3 4 MODEON Print at receiving data.Standard mode, statistical calculation modeMODE 21 2 3 4 MODEON Print by the AD-8121B DATA key operation or AD-8121B built-in timer.Standard mode, interval mode, chart modeMODE 31 2 3 4 MODEON Print at receiving data.Dump print modeDIP switch No.3 : Handling unstable dataON PrintOFF Not printedSet the DIP switch No.4 to OFF.1 2 3 4 MODEON1 2 3 4 MODEON