489-10 Clock and Calendar FunctionThe balance is equipped with a clock and calendar function. When the “GLP output (info)” parameter isset to “1” or “2” and the “Time/Date output (5-td)” parameter is set to “1”, “2” or “3”, the time and date areadded to the output data. Set or confirm the time and date as follows:Operation1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.2 Press the SAMPLE key several times to display Cl adj .3 Press the PRINT key. The balance enters the mode to confirm or set the time and date.Confirming the date4 The current date is displayed with all the digits blinking.When the date is correct and the operation is to be finished, press the CAL key and proceed to step 8.When the time is to be confirmed again, press the SAMPLE key and go back to step 4.When the date is not correct and is to be changed, press the RE-ZERO key and proceed to step 7.NoteThe year is expressed using a two-digit format. For example, the year 2000 is expressed as “00”.Setting the date (with part of the digits blinking)5 Set the Year, Month and Date using the following keys.SAMPLE key To select the digits to change the value.The selected digits blink.RE-ZERO key To increase the value by one.MODE key To decrease the value by one.PRINT key To store the new setting, displayend and go to step 6.CAL key To cancel the new setting and go to step 6.Confirming the time6 The current time is displayed with all the digits blinking.When the time is correct and the date does not need to beconfirmed, press the CAL key and proceed to step 8.When the time is correct and the date is to be confirmed,press the SAMPLE key and proceed to step 4.When the time is not correct and is to be changed,press the RE-ZERO key and proceed to step 7.Setting the time (with part of the digits blinking)7 Set the Hour and Minute in 24-hour format using the following keys.SAMPLE key To select the digits to change the value.The selected digits blink.RE-ZERO key To increase the value by one.MODE key To decrease the value by one.PRINT key To store the new setting, displayend and go to step 8.CAL key To cancel the new setting and go to step 8.ConfirmIn data setting modeTo Hour & MinuteMonth Date YearTo Month Date YearConfirmHour & Minute