CHAPTER 4: Troubleshooting234.6 Heart Rate Function IssuesSymptom: Heart Rate Function Does Not Work or is Reading IncorrectlySolution:1. With a multi-meter set for DC voltage, place one terminal on each of the HR grip plates. The HR gripshould give a voltage reading of between 0.5 and 2.0VDC. If the voltage is not between 0.5 and 2.0VDC,remove the 3 screws holding the HR grip together and check the connection of the HR grip wiring.2. Check continuity of the HR grip wiring.Place one terminal of a multi-meter set for resistance on the HR grip wiring at the HR grip, and theother terminal on the HR grip wiring at the console. An ohm reading of around 1 should be expected.If the reading is higher than 1, replace the HR grip wiring.3. Remove the console. Remove the 6 screws holding the front of the console to the rear. Check theconnection of the HR board wiring to the UCB. If all the wiring is intact and has good contact, replace theHR board.4.If the HR board, HR grips, and HR grip wiring do not solve the issue, replace the UCB.