Parameterization92 FSM4000 D184B140U027.4.2 Recommended settings for diagnostic limit valuesThe values indicated are only intended as a rough guide and may need to be adapted in linewith on-site conditions.Parameter Minimum value Maximum valueElectrode voltage 1 These values depend on the flowrate, so no recommendations can be made here.Electrode voltage 2Electrode balance 1) -150 % 150 %Coil current Factory or commissioning fingerprint x 0.95 Factory or commissioning fingerprint x 1.05Coil voltage This value depends on the fluid temperature Tmedium .Tmedium Tmedium-40 °C(-40 °F) Factory fingerprint x 0.81 -40 °C(-40 °F) Factory fingerprint x 0.89-20 °C(-4 °F) Factory fingerprint x 0.86 -20 °C(-4 °F) Factory fingerprint x 0.950 °C(32 °F) Factory fingerprint x 0.9 0 °C(32 °F) Factory fingerprint x 120 °C(68 °F) Factory fingerprint x 0.95 20 °C(68 °F) Factory fingerprint x 1.0560 °C(140 °F) Factory fingerprint x 1.05 60 °C(140 °F) Factory fingerprint x 1.1690 °C(194 °F) Factory fingerprint x 1.14 90 °C(194 °F) Factory fingerprint x 1.26130 °C(266 °F) Factory fingerprint x 1.24 130 °C(266 °F) Factory fingerprint x 1.37180 °C(356 °F) Factory fingerprint x 1.33 180 °C(356 °F) Factory fingerprint x 1.47Coil resistance This value depends on the fluid temperature Tmedium .Tmedium Tmedium-40 °C(-40 °F) Factory fingerprint * 0.71 -40 °C(-40 °F) Factory fingerprint * 0.79-20 °C(-4 °F) Factory fingerprint * 0.81 -20 °C(-4 °F) Factory fingerprint * 0.890 °C(32 °F) Factory fingerprint * 0.9 0 °C(32 °F) Factory fingerprint * 120 °C(68 °F) Factory fingerprint * 0.95 20 °C(68 °F) Factory fingerprint * 1.0560 °C(140 °F) Factory fingerprint * 1.19 60 °C(140 °F) Factory fingerprint * 1.3190 °C(194 °F) Factory fingerprint * 1.28 90 °C(194 °F) Factory fingerprint * 1.42130 °C(266 °F) Factory fingerprint * 1.43 130 °C(266 °F) Factory fingerprint * 1.58180 °C(356 °F) Factory fingerprint * 1.62 180 °C(356 °F) Factory fingerprint * 1.79Coil temperature This value depends on the ambient / fluid temperature, so no recommendations can bemade here.1) This function is only guaranteed if an electrode zero point adjustment has been carried out (see page 85).