CommissioningD184B140U02 FSM4000 496.2 Commissioning the unit6.2.1 Switching on supply powerAfter switching on the supply power, the flowmeter sensor data in the external FRAM iscompared with the data saved internally. If the data is not identical, the transmitter data isreplaced automatically. Once completed, Warning 7 "Flowmeter sensor data loaded" andWarning 8b "Update external FRAM" are displayed. The measuring equipment is now ready foroperation.The display shows the current flowrate.6.2.2 Device configurationThe device can be configured at factory to customer specifications upon request. If no customerinformation is available, the device is delivered with factory settings.On-site configuration requires only a few parameter settings. For information on settings, referto the "Entering data in short form". A short overview of the menu structure can be found in thesection "Parameter overview".The Easy Set-up menu enables users to configure the unit quickly and conveniently, seechapter "6.3 Easy Set-up: For uncomplicated configuration".The following parameters should be checked or set before start-up:1. Flow range end value (menu items “Q max” and “Unit”).The devices factory setting is largest flow range end value, unless other customerinformation is available. The ideal flow range end values are approximately 2 ... 3 m/s. Firstset the unit Qmax (e.g., m 3 /h or l/s) under menu item “Unit”, and then set the flow range endvalue under “Qmax”. The smallest and largest possible flow range end values are shown intable "Flowmeter Sizes, Pressure Ratings, Flow Range" page 29.2. Current output (menu item “Current output”)Select the desired current range (0 … 20 mA or 4 … 20 mA)3. For devices with a fieldbus, the bus address must be set (menu item “Data Link”).4.Pulse output (menu items “Pulse” and “Unit”).To set the number of pulses per volume flow unit, a unit for the totalizer (e.g., m 3 or l) mustbe selected under “Unit”. Afterward the number of pulses has to be entered in the menu item“Pulse output”.5. Pulse width (menu item “Pulse output”)For external processing of the counting pulses, the pulse width at terminals 51 and 52 canbe set between 0.1 ms and 2.000 ms.