Page 87025 Instruction Manual (Rev. 2)4.7 Connection to AnimalA “Y” connection should be inserted between the tracheal cannula and the two tubescoming from the pump.Dead space is minimized if the “Y” is as close to the trachea as possible. See Figure 5“Valve Connection to Animal”.Expired air can be recovered via the exhaust tube for sampling, partial recycling, etc.,see Figure 5 “Valve Connection to Animal”.Figure 5 “Valve Connection to Animal”Room air or any anaesthetic gas mixture can be used to feed the pump intake.5 MAINTENANCEWhile any service of the instrument ought to be carried out by Ugo Basile personnel or byqualified personnel authorized by UGO BASILE organization, this section of the instruc-tion manuals describes normal maintenance procedures which can be carried out at yourfacility.UNPLUG THE MAINS CORD BEFORE CARRYING OUTANY MAINTENANCE JOB!5.1 ElectricalTo inspect and/or replace the fuses, disconnect the mains cable first! Insert a minia-ture screwdriver in the slot indentation, see paragraph 3.3-Fuse Holder, and snap out theslide which houses the fuses. We recommend 800mA timed fuses, type T800.After replacing the fuses, snap in the fuse slide: the mechanical “click” ensures that it islocked.