47PBD-51041064 SITRANS LG270 - Operating Instructions44293-EN-171021Menu - Additional adjustmentsMenu Menu item Default valueAdditional adjustments PIN 0000Date Actual dateTime Actual timeTime - Format 24 hoursProbe type Device-specificThe instrument settings are copied with this function. The followingfunctions are available:• Read from sensor: Read data from sensor and save in the displayand adjustment module• Write to sensor: Save data from the display and adjustment mod-ule back into the sensorThe following data or settings for adjustment of the display and ad-justment module are saved:• All data of the menu "Setup" and "Display"• In the menu "Additional adjustments" the items "Reset, Date/Time"• Special parametersThe copied data are permanently saved in an EEPROM memory inthe display and adjustment module and remain there even in case ofpower failure. From there, they can be written into one or more sen-sors or kept as backup for a possible electronics exchange.Note:Before the data are stored in the sensor, a check is carried out todetermine if the data fit the sensor. If the data do not fit, a fault signalis triggered or the function is blocked. When data are being writteninto the sensor, the display shows which instrument type the dataoriginate from and which TAG-no. this sensor had.Tip:We recommend to save the instrument adjustments. In case of anelectronics exchange the saved parameter adjustment data relievethis process.Since scaling is very extensive, scaling of the level value was dividedinto two menu items.Additional settings - Copyinstrument settingsAdditional settings - Scal-ing level