Connecting4.1 Electrical connectionsQuick Start18 Compact Operating Instructions, 01/2013, A5E31805656-014.1.1.2 Wall-mounted deviceWARNINGCable glandsThe cables must have the correct diameter for the cable glands. After installation, makesure that the cables are positioned securely and that the seals are in the correct position.WARNINGDanger due to electrical currentCheck that the existing power supply corresponds to that specified on the nameplate of thedevice.WARNINGDanger due to electrical currentThe power supply cable to be connected at the customer location could be damaged byincorrect fitting in the device.You should therefore carry out a visual inspection and PE conductor test of the cable inaccordance with the country-specific safety standards, regulations or other applicabledirectives before starting up the device. Since this is a cable which lies partially within thedevice, you must always additionally apply the production safety standard IEC / EN 61010-1 or its national version.4.1.2 Rack-mounted device4.1.2.1 Electrical connection of the power supplyProcedureNoteAppliance plugYou can use any appliance plug that fulfills the conditions of DIN EN 60320-1, ≥6 A.The mains supply line must comply with the valid regulations and conditions for the place ofinstallation and be provided with a protective conductor at the potential of the housing. Thecross-section of each wire has to be ≥ 1 mm2. The phased terminal lead must be connectedin the connector at the position marked.