Climatic conditionsAmbient humidity (rel. humidity)during storage, transport or opera‐tion< 90%Permissible ambient temperatureand humidityTechnical specifications of analyzer modules (Page 163)13.3 Technical specifications of analyzer modules13.3.1 OXYMAT 7 technical specificationsTable 13-7 OXYMAT 7: General technical specificationsGeneral OXYMAT 7 OXYMAT 7 high-tempera‐ture 1)Information on the smallest span See module nameplate or device certificateMax. power consumption Warm-up phase ≤ 60 W ≤ 120 WWeight (only module) < 5.5 kg < 7.5 kgElectrical safety In accordance with DIN EN 61010-1Max. EMC error during operation with one and/or twoOXYMAT 7 / ULTRAMAT 7 units1.0% according to EN61326-1 (2013)1) The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.Table 13-8 OXYMAT 7: Measuring rangesMeasuring ranges OXYMAT 7 OXYMAT 7 high-temperature 1)Number of measuring ranges Max. 4; parameters can be as‐signed freelyMax. 4; parameters can be as‐signed freelyParameters can be as‐signed in the measur‐ing rangesSmallest possible measuringspan0.5% O2 2% O2Largest possible span 100% O2 100% O21) The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.NoteTemperature change for high-temperature OXYMAT 7If you change the temperature range, the device has to be re-calibrated. Detailed informationcan be found in the operating manuals, see "Bibliography" (Page 179).Technical specifications13.3 Technical specifications of analyzer modulesWall-mounted deviceOperating Instructions, 07/2023, A5E31930403-AB 163