Checking pressure switch for function1. Increase the reference gas pressure until the pressure switch is triggered, meaning that theswitch contact opens.2. Please ensure that the reference gas pressure is not more than 2 000 hPa above the samplegas pressure.The value when pressure is increasing (response pressure) is always higher than the valuewhen pressure is decreasing (release pressure). This so-called switching hysteresis of thepressure switch is approx. 200 hPa.9.5 Zero point errorNoteZero point error because of surrounding gasesKeep the ambient air of the analyzer section free from the measured gas component; otherwise,measurement results will be affected. Gases which exhibit cross-sensitivity to the measured gascomponent must also not be present in the ambient air.9.5.1 OXYMAT 7Accompanying gases (interfering gases) cause a zero point deviation that is positive withparamagnetic gases and negative with diamagnetic gases. The extent of this offset per 100 vol.% accompanying gas, expressed in vol. % O2 absolute, is listed in the following table. The valuesin the tables refer to nitrogen at 60 °C and 1 000 hPa absolute according to IEC 61207/3.Table 9-3 Zero point error of organic gasesAccompanying gas (concentration 100 vol. %) Zero point deviation invol. % O2 absoluteEthane C2H6 -0.49Ethene (ethylene) C2H4 -0.22Ethine (acetylene) C2H2 -0.291.2 butadiene C4H6 -0.651.3 butadiene C4H6 -0.49n-butane C4H10 -1.26Isobutane C4H10 -1.301-butene C4H8 -0.96Isobutene C4H8 -1.06Dichlorodifluoromethane (R12) CCl2F2 -1.32Ethanoic acid CH3COOH -0.64n-heptane C7H16 -2.40n-hexane C6H14 -2.02Cyclo-hexane C6H12 -1.84Commissioning9.5 Zero point errorWall-mounted device130 Operating Instructions, 07/2023, A5E31930403-AB