- 54 -21.4.16. Send all implemented commandsFormat: PC CR LFResponse: PC_- >_Z,T,OT,UT,S,SI,SU,SUI,C1,C0,CU1,CU0,K1,K0,NB,PC -command carried out, all implemented commands have been sent.21.5. Manual Printout / Automatic PrintoutIt is possible to generate printouts either manually or automatically.• Manual printout is generated for stable weighing result. Load the platform,wait for a stable result and press key.• Automatic printout is generated for stable weighing result. Load theplatform, wait for a stable result. No key needs to be pressed.Caution:For verified balance option of temporary weighing results printout is disabled.Format:1 2 3 4 -12 13 14 15 16 17 18stabilitymarker space character mass space unit CR LFStability marker [space] if measurement result stable[?] if measurement result unstable[^] if high limit is out of range[v] if low limit is out of rangecharacter [space] for positive values[-] for negative valuesMass 9 characters with decimal point, right justificationunit 3 characters, left justificationCommand 3 characters, left justificationExample 1:_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - printout generated upon pressingENTER/PRINT key.Example 2:? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - printout generated upon pressingENTER/PRINT key.