EN-38 PR Series Balance9.3 AccessoriesDESCRIPTION PART NUMBERAuxiliary Display AD7-RS 30472064Security Device 80850043RS232 Cableδ25 pinε 80500524RS232 Cableδ9 pinε 80500525Dust Cover 30093334In-use Cover 30372547Printer SF40A 30064202 (EU); 30064203 (AM)Power Adapter for Balance 460017249.4 Communication9.4.1 Interface CommandsCommands listed in the following table will be acknowledged by the balance.Command FunctionIP Immediate Print of displayed weight (stable or unstable).P Print displayed weight (stable or unstable).CP Continuous Print.SP Print on Stability.SLP Set Auto Print to On Stability, allow non-zero displayed weight be printed.SLZP Set Auto Print to On Stability, allow both stable non-zero weight and stable zero reading to beprinted.xP Set Auto Print to Interval Print, x = print interval (1-3600 sec), 0P disable the interval Print0P 0P disable interval print, continuous print or print on stabilityH Enter Print Header Lines, the format is: H x "header string".Where x = line number 1 to 3, “header string” can be up to 24 alphanumeric characters.If no string in the command, "H x" will read the stored header x.Z Same as pressing Zero Key.T Same as pressing Tare Key.xT*** Establish a preset Tare value in displayed unit. x = preset tare value. Sending 0T clears tare (ifallowed).PT Prints Tare weight stored in memory.PM Print current application mode (weighing mode).xM Set current application mode to x. x depends on applications1 – Weigh2 – Count3 – PercentM Scroll to the next enabled mode.PU Print Current weighing unit: g, kg, lb, oz, etc….xU Set balance to unit x: g, kg etc.1 – g2 – kg3 – mg4 – ct