CALIBRATION PROCEDUREPXIe-1095PXI ChassisThis document contains the verification and adjustment procedures for the NationalInstruments PXIe-1095 chassis. Refer to ni.com/calibration for more information aboutcalibration solutions.ContentsSoftware.................................................................................................................................... 1Documentation..........................................................................................................................1Test Equipment......................................................................................................................... 2Test Conditions......................................................................................................................... 2Initial Setup...............................................................................................................................2Verification................................................................................................................................3Adjustment................................................................................................................................4Reverification............................................................................................................................5Where to Go for Support.......................................................................................................... 5SoftwareCalibrating the PXIe-1095 requires the installation of the following software on the calibrationsystem:• PXI Platform Services 18.1 or laterYou can download PXI Platform Services from ni.com/downloads.PXI Platform Services supports programming the calibration procedures using a number ofapplication development environments (ADEs), including LabVIEW, LabWindows™/CVI™,and Microsoft Visual C++. When you install PXI Platform Services, you need to installsupport only for the ADE that you intend to use.DocumentationYou might find the following documents helpful as you perform the calibration procedure:• PXIe-1095 User Manual• PXIe-1095 SpecificationsThe latest versions of these documents are available on http://www.ni.com/manuals.