CALIBRATION PROCEDUREPXIe-1092PXI ChassisThis document contains the verification and adjustment procedures for the NationalInstruments PXIe-1092 chassis. Refer to ni.com/calibration for more information aboutcalibration solutions.ContentsSoftware.................................................................................................................................... 1Documentation.......................................................................................................................... 1Test Equipment..........................................................................................................................2Test Conditions..........................................................................................................................2Initial Setup............................................................................................................................... 3Verification................................................................................................................................3Adjustment................................................................................................................................ 4Reverification............................................................................................................................6Where to Go for Support...........................................................................................................6SoftwareCalibrating the PXIe-1092 requires installing the following software on the calibration system:• PXI Platform Services 18.1 or laterYou can download PXI Platform Services from ni.com/downloads.PXI Platform Services supports programming the calibration procedures using severalapplication development environments (ADEs), including LabVIEW, LabWindows™/CVI™,and Microsoft Visual C++. When you install PXI Platform Services, you need to installsupport for only the ADE that you intend to use.DocumentationYou might find the following documents helpful as you perform the calibration procedure:• PXIe-1092 User Manual• PXIe-1092 SpecificationsThe latest versions of these documents are available on ni.com/manuals.