10.6.1 Temporal synopticVERY IMPORTANT: when audio function is activated, the DI duration of elementary integration for calculation ofthe LAeq et LCeq equivalent continuous level is automatically set at 1 second.10.6.2 Synoptic of the management of the recording dynamicSelection of a detection level for audio recording: it must be on the planned measuring range: 40-120 or 60-140 in two-channel operating V1 and V2 or 40-140 in one-channel operating V1.Example: do not choose a 125 dB threshold on the 40-120 range !!!24Final recording: pre and post exceeding + exceeding707580859095100105LAeq DISeuildB1401201008060400 dB24dB48dBSeuil enreg.Dynamique de mesurede V1 ou V1/V2Gain numérique pourenregistreurniveau d'écouteconfortableniveau d'écoutemoyen à faibleTgl: pre and post recording fixed at 4 seconds for supervision of audio exceeding on thresholdExceeding area of sound level on threshold: ex : 87 dB