Keysight E5063A Network Analyzer 21InstallationPower SupplyPower SupplyBefore turning on the E5063A power, check the following.Verification of the Power SupplyConfirm that the power supplied to the E5063A meets the followingrequirements:Verification and Connection of Power CableThe three-wire power cable attached to the E5063A has one wire serving as aground. Using this power cable allows the E5063A to be grounded, therebyprotecting you against electrical shock from the power outlet.Step 1. Confirm that the power cable is not damaged.Step 2. Use the supplied cable to connect between the power cable receptacle(Figure 1-3 on page 24) on the rear panel of the E5063A and a three-wirepower outlet with the grounding prong firmly connected in the ground slotPower cord list, 16000-99101 shows the power cable options.RequirementsVoltage 90 to 132 VAC or 198 to 264 VAC **. Switched automatically by the E5063A in conformity to the voltage used.Frequency 47 to 63 HzMaximum power consumption 300 VANEVER use a power cable showing any sign of damage. Faulty cables cancause electrical shock.Use the supplied power cable with grounding wire to securely ground theE5063A.