18Specifications§5Product Handheld Particle CounterModel 3888 (3-channel model)3889 (6-channel model)ParticleMeasurementMeasuringmethod Light scatteringSize distribution 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm)3 channels (0.3, 0.5, 5.0μm)Flow rate0.1CFM (2.83L/min)Accuracy: ±5% (Compliant with JIS B9921 andISO21501-4)Sampling time From 6 seconds to 99 minutes and 59 seconds(for 1 measurement cycle)Interval timeFrom 6 seconds to 99 minutes 59 seconds(Interval between the start and the next start of the repeatmeasurements)Sampling cycle From 1 to 999 cycles or continuous measurementsLocationclassification 99 locationsCalibration NIST traceableMeasuring mode Repeat, Single, Continuous, Calculation, ISO, GBDisplay time ofmeasured value From 1 to 10 secondsDisplay ofmeasured value Differential Δ and Cumulative∑MaximummeasurableConcentration2,000,000 particles/CF at 10% coincidence loss(compliant with JIS B9921 and ISO21501-4)Countingefficiency50±20%(for PSL particles near the minimum measurablesize)100±10%(for PSL particles of 1.5 to 2 times as large as theminimum measurable size)(Compliant with JIS B9921 and ISO21501-4)False count ≦1 particle/5 minutes(Compliant with JIS B9921 and ISO21501-4)Size resolution ≦15% (for PSL particles near 0.3μm)(Compliant with JIS B9921 and ISO21501-4)Pump Internal pump(Vane)Exhaust With filterDisplay 4.3 inch color LCD, Resistive touch panel