Supplied By www.heating spares.co Tel. 0161 620 6677KANE450 manual Page 21MEASURING AMBIENT AIR (ROOM CO)The analyser's CO sensor can be used to detected spillage from appliances. A standardspillage test can be conducted where the analyser automatically records 16 sets of 1 minutesamples and prints the results as below:Standard Room CO Report .... this mode is accessed via the MENU functionK450 1.0Serial No. 123456789sales@kane.co.ukKANE INTERNATIONAL..............................................Room CO Test..............................................LOG 05TIME 16:58 12/12/05TEST CO ppm0 11 52 73 94 105 136 167 248 309 2810 1811 1412 1013 614 415 3........................................MAXIMUM CO 30.........................................Customer.........................................Appliance..........................................Ref:...........................................