Option DescriptionOptions Sets the following options:• Set the application parameters view to expert or basic mode• When the instrument is set to off, sets the syringe to empty into thetitrant bottle• Changes the temperature display unit• Prints the measurement and derivative curves if a printer is connected• Specify if a balance is connected• Specify if an external propeller stirrer is connected.Note: When the instrument is set to on, the application parameters view isalways set to basic mode. Changes made to application data with theinstrument in expert mode are kept.Security Change the password. Specify which of these options are passwordprotected:• Settings (all items except Info)• Maintenance schedule and summary• Delete data log• Export data log• Electrode calibration• Titrant calibration• Sample volume calibration• Syringe verification test3. Push Back.6.2 Install the applicationsUse the supplied USB key to install the applications. The instrument can install amaximum of five applications for each syringe installed. For two syringes, the installedapplications shown on the top line of the display refer to syringe one and the installedapplications shown on the second line refer to syringe two.If any errors occur during installation, refer to Troubleshooting on page 49.1. Push Home to go to the main menu.2. Connect the USB key to the USB port on the side of the instrument. The applicationson the USB key show on the display.3. Push the arrow keys to highlight and select an application to install. Push the left orright arrow key to select it. Do this step again to select additional applications toinstall.4. Push Import to install the selected applications.5. Push OK to complete the installation. The installed applications show on the homescreen.Note: To install more applications, push Home to go to the home screen, then remove theUSB key and reconnect it.6.3 Prepare the instrument for measurement1. From the home screen, select Purge. All attached devices are listed.2. Select All elements to purge all the attached devices, or select one device to purge.Push Select. Air is removed from the device and filled with liquid from the bottle.3. Push OK when the operation has completed.4. Make sure that there are no air bubbles in the device. Do step 2 again if there are anyair bubbles.Startup30