-1-Cellular Antenna Quick Start GuideHP 8712ET/ES and HP 8714ET/ESHP Part Number 08714-90022Printed in USA October 1998© Copyright 1998 Hewlett-Packard CompanyThis quick start guide provides basic instructions on how to verify the performance of cellular antennasystems. Refer to the Option 100 Fault Location and Structural Return Loss Measurement User’s GuideSupplement for more detailed information. Also, please refer to your analyzer’s User’s Guide for safety,warranty, and assistance information.NOTE This quick start guide assumes the use of an HP 8712ET or HP 8714ET. If you are using an HP 8712ES orHP 8714ES, some key presses and displays will be slightly different.To Make a Fault Location Measurement1. Choose the measurement parameters or recall an instrument state.2. Reduce the interference. 3. Calibrate the analyzer.PRESET BEGIN Cable Fault LocationStart DistanceXX ENTER Stop DistanceXX ENTERBEGIN Cable Fault LocationBand Pass Center Frequency 900MHzCAL Full Band CalFREQ Fault Loc FrequencyBand Pass Max Span 300 MHzAVG System BandwidthNarrow 250 Hz