Canon CXDI-31 Service Manual
2 Installation Manual- 100 -(3) Startup method1) Start up Windows NT.2) Launch “ccrstart.bat” which is located in “D:\ccr”*When the CXDI application is launched, the message Sensor not connectedappears. Click [OK].3) When the exposure screen appears on the operation unit, press [Alt] + [Tab] on thekeyboard to enter debugging mode.4) When Welcome to CCR appears, select “1. Set-Up...”.5) When Setting Mode (0: Normal, 1: Expert) [0=0x0]: appears, select “1: Expert.”6) When CCR SETUP MENU appears, select “7 Scan Sensor Setup”.7) When Capture Device Configuration Table appears, enter dummy serial numbers for theitems indicated below.The dummy serial numbers are as follows:Upright stand Serial Number 00000199Patient table Serial Number 00002009Universal Serial Number 0000123Cassette Serial Number 0000302* Be sure to write down the correct sensor serial numbers before entering thedummy serial numbers.For example, when the upright stand model is connected to Sensor 1, and the table modelis connected to Sensor 2:Enter “0000199” for “A/D Board Serial Number for Sensor ID#1”Enter “00002009” for “A/D Board Serial Number for Sensor ID#2”The CXDI application can now be launched on the control PC with the same conditionsin effect as if an imaging unit were connected. |