R5911227 /02 Coronis Fusion 4MP / 6MP 294.17 Continuous ALCAbout Continuous ALCContinuous ALC can only be selected when the DICOM display function is selected. Therefore,please refer to “Display functions”, page 27 to correctly set the display function.Enabling continuous ALC will continuously recalculate the DICOM display function taking the averagedambient light into account.To select continuous ALC:1. Bring up the OSD main menu.2. Navigate to the Configuration > Calibration > Ambient Light menu.3. Enter the Continuous ALC submenu.4. Select Enabled/Disabled as desired and confirm.4.18 Embedded QAOverview• About Embedded QA• DICOM status report• DICOM compliance check• DICOM calibration• Reset DICOM calibration• DICOM error threshold4.18.1 About Embedded QAAboutEmbedded QA allows you to run a display calibration or compliance test directly from the display using theOSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure thenecessary luminance levels for either a calibration or compliance test. Various settings for both actions can beselected from the display’s OSD menu. The last results of both actions can be consulted from the OSD.Embedded QA or MediCal QAWeb?Embedded QA is not a replacement for the Barco MediCal QAWeb solution.Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco stillhighly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb bringsmany benefits such as centralized asset management, the ability to schedule tasks, remote management,automated reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA andAAPM TG18. That's why MediCal QAWeb Agent acts as the master for all supported displays from themoment it is installed and running. MediCal QAWeb Agent will take over from Embedded QA and overwriteany settings which were applied by Embedded QA.4.18.2 DICOM status reportAbout DICOM status reportFollowing information is available:DICOM Compliance Status (status since last compliance check)• Compliance status: Shows if the current DICOM curve is compliant or not.• Maximum error: Shows the maximum error of the current DICOM curve. This is the deviation compared toa perfect DICOM.Advanced operation