6. Tap the nozzle on a hard surface to remove any last remaining particles.7. Rinse the nozzle under water.8. Rinse the body under water.9. Use a syringe to rinse the parts that are hard to reach.10. Dry the nozzle and body using a soft, lint-free disposable cloth.11. Run the nozzle and body through the washer-disinfector.6.4 SterilizationThe disposable sterilization pouches must comply with ISO 11 607 or any equivalent standard required bynational regulations.1. Remove the nozzle and body from the washer-disinfector.2. Dry them.3. Package them in a sterilization pouch, with each item in its own pouch.4. Sterilize them in the steam autoclave with air space according to the normal cycle in your activity area.In Europe, depending on the country:l 18 minutes at 134°C and 20 minutes of dryingl 4 minutes at 134°C and 20 minutes of dryingl 3 minutes at 134°C and 20 minutes of dryingA minimum pressure of 2 bars.In the U.S.A. - 4 minutes at 132 °C with 20 minutes of dryingA minimum pressure of 1.85 bars.6.5 StorageStore the sterile items at room temperature in a dry location, away from dust. Prior to use, check the integrity of thepackaging, and if necessary, repeat the sterilization.If there is visible contamination in the bag, place the item in an infectious medical waste container to be safelydestroyed.Page 16/35 - User Manual • AIR-N-GO® easy • J10129 • V5 • 11 • 10/2015 • ND27US050E