Disinfecting a Device and Accessories for Use by Multiple ClientsNote: This section applies only when multiple clients will be using the same device. In thatsituation, it’s vital to effectively disinfect the device and accessories that have been used by aclient before handling and use by another client.First Step: Put on protective glovesNext Step: Clean the device and accessories1. Wipe down the device and accessories with disinfectant wipes (Virucidal, Bactericidal,Pseduomonicidal, Tuberculocidal, Fungicidal)—Metrex CaviWipes™. Follow the disinfectant productmanufacturer’s instructions for cleaning.2. Take a Q-tip with cleaner applied (for example, Windex® or other commercial cleaner) and cleaninside all cracks and crevices. You may want to use a toothbrush or similar brush.3. Blow the device off with an air hose, or wipe it dry.Important! Make sure all soil is removed from the device and accessories. This is vitalbefore proceeding to disinfecting the equipment.Next Step: Disinfect the device and accessories1. Wipe down the device and accessories with a new disinfectant wipe (Virucidal, Bactericidal,Pseudomonicidal, Tuberculocidal, Fungicidal)—Metrex CaviWipes. Follow the disinfectant productmanufacturer’s instructions for disinfecting.2. Make sure to clean inside all cracks and corners, and wipe more than once if necessary to keep thedevice wet for a minimum of three (3) minutes.3. Allow the device and accessories to dry.Final Step: Wipe the touch screen1. When the device and accessories are dry, wipe down the viewing area (device touch screen,TouchGuide, Keyguard, etc.) with glass cleaner, so the screen doesn’t discolor.2. Allow the equipment to dry.Accent 1000 Hardware Manual 31