95. OPERATION5-1. Turning the power ON and OFF1. Press the ON/OFF key to turn the power ON.All the display symbols appear as shown above.(As for weighing units: Only the units stored in memory are displayed.)The display turns off except a weighing unit and a decimal point.The scale waits for the weight value to become stable, and zero is displayed withthe ZERO indicator (power-on zero).The range for power-on zero is within ±10% of the weighing capacity around thecalibrated zero point.If the power is turned ON while there is a load beyond this range, the scale is taredto zero and the NET and ZERO indicators turn on.2. Press the ON/OFF key to turn the power OFF. Auto power-off functionThe auto power-off function automatically turns the power OFF when zerois displayed for approximately 5 minutes.See “poff” in “8-5. Function list.” Display lock functionThe display lock function locks the display temporarily to make the weightvalue easier to read.See “Hold” in “8-5. Function list.”LCD backlight functionThe LCD backlight turns on when the weight value changes 4d or more (d= minimum display, so, corresponding to four times as great as theminimum display) or any key operation is performed. When the weightvalue is stable for a certain period of time, the backlight automaticallyturns off. The backlight can also be set to remain on or off.See “ltUp” in “8-5. Function list.”