208-5. Function listClass Item Param-eter Description 0 Auto power-off disabledpoffAuto power-off 1 Auto power-off enabledAutomaticallypower off0 Fast / sensitive1 23CondResponse4 Slow / stableSoftware filtering0 Stable when within ± 0.5d/0.5s 1 Stable when within ± 1d/0.5sst-bStability bandwidth 2 Stable when within ± 2d/0.5sConditions toturn on theSTABLE indicator0 DisabledtrcZero tracking 1 EnabledTracking zero shift 0 Point (.)pntDecimal point 1 Comma (,)Decimal separator 0 Comparator disabled1 Compares all data2 Compares all stable data3 Compares plus data > +4d4 Compare stable plus data > +4d5 Compares data > +4d or < -4dCpComparator mode6 Compares stable data > +4d or < -4dConditions tocompare.d = the minimumdisplay division0 Command and stream modes 1 Command and PRINT key2 Command, PRINT key and auto-print A3 Command, PRINT key and auto-print BprtData outputmode4 Command mode onlyAuto-print A:+ dataAuto-print B:+/- data 0 No pause (general equipment)pUseData output pause 1 1.6 seconds (for AD-8121)Interval betweencontinuous data 0 No output1 AD-8121 format (* )infoGLP output2 General formatGLPoutput format 0 2400 bps1 4800 bps2 9600 bpsbpsBaud rate3 1200 bps 0 7 bits, even parity1 7 bits, odd parityfuncbtprData and parity2 8 bits, non parity Factory setting(* ) When the AD-8121 format is selected, the interval between data is 1.6 secondsregardless of the setting “pUse”.