10.2 Battery PacksUse only Defibtech battery packs in the DDU-2000 Series AED.DBP-2003 and DBP-2013 Battery PacksCategory SpecificationModel number DBP-2003DBP-2013 (Aviation;TSO C-142a)Main battery type 12VDC, 2800 mAh, Lithium/Manganese Dioxide.Disposable, recyclable, non-rechargeable.Capacity 125 shocks or 8 hours of continuous operation.*Charge time 4 seconds or less (from shock advised)*Standby-life (installed in unit) 4 years**Typical, new battery, at 25°CNote: For information about optional rechargeable lithium-ion battery packs, please visit www.defibtech.com orcontact Defibtech or your distributor.10.3 Self-Adhesive Defibrillation PadsDefibtech self-adhesive defibrillation pads have the following characteristics:Model number DDP-2001 DDP-2002Type Adult Child/Infant < 8 years< 55 lbs. (25 kg)Intended use Disposable DisposableAdhesion Self-adhesive Self-adhesiveActive gel surface area 12 inches2 (77 cm2) each (nominal) 7.75 inches2 (50 cm2) each (nominal)Cable/connector type Integrated IntegratedCable length 4 feet (122 cm) (typical) 4 feet (122 cm) (typical)Expiration date 2.5 years from date of manufacture 2.5 years from date of manufacture10.4 ECG Monitoring Adapter / Cable (optional)For use only with DDU-2450 AEDs.Model number DAC-2020 and DAC-2021Patient connection Type CF, fully defibrillation-protectedCable length 6.5 feet (2 meters)Patient lead wire designationLead type DAC-2020 (AAMI) DAC-2021 (IEC) PlacementPositive Red - LL Green - F Left legNegative White - RA Red - R Right armReference Black - LA Yellow - L Left armTypical lead connection Lead IIPerformance withDDU-2450 AED Meets environmental specifications for DDU-2000 Series.10. Technical Specifications56DAC-A2510EN-BG