R3765/67G Series Network Analyzer Operation Manual10.1 Preparing for a Performance Test10-3Table 10-1 Required Measurement Instrument for Performance Test (3 of 3)10.1.3 General Note• Use an AC power source having a voltage of 90V to 250V and a frequency of 48Hz to 66Hz.• When connecting the power supply cable, turn OFF the POWER switch.• The R3765/67G series must be tested under the following conditions:Temperature : +23°C to ±5°CRelative humidity : 80% RH or lessFree from dust, vibration and noise.Test items Measurement Instrument RemarksDynamicaccuracy• Step attenuatorVariable range: 0 dB to 90 dBAccuracy : Within 0.02 dBHP8496B(Calibrated under thenational standard)Refer tosection 10.9• RF cables(SMA(m)/SMA(m) 50Ω) × 2A01253-060• Conversion connectors(N(m)/SMA(f)) × 2HRM-554S• 3dB fixed attenuators(SMA(f)/SMA(m)) × 2AT-103• 50Ω to 75Ω conversion adapter(when OPT 12 or 13 is installed)(75ΩN(m)/50ΩN(f)) × 211852B(Agilent Technolo-gies)Attenuationaccuracy(OPT10)• RF cable(SMA(m)/SMA(m) 50Ω)A01253-060• Conversion connectors(N(m)/SMA(f)) × 2HRN-554S• RF cable (For OPT12 or 13)(N(m)/N(m) 75Ω)8120-6468(Agilent Technolo-gies)