__________________________________________________________________________ EnglishVAISALA _______________________________________________________________________ 15Technical DataRelative Humidity Measurement SpecificationsProperty Description / ValueMeasurement range 0 ... 100 % RHAccuracy (including non-linearity,hysteresis, and repeatability):at 0 ... +40 °C (+32 ... +104 °F) ±1.7 %RH (0 ... 90 % RH)±2.5 %RH (90 ... 100 % RH)at -40 ... 0 °C and +40 ... +80 °C(-40 ... +32 °F and +104...+176 °F) ±3.0 %RH (0 ... 90 % RH)±4.0 %RH (90 ... 100 % RH)Factory calibration uncertainty at20 °C (+68 °F) ±1.5 %RHHumidity sensor HUMICAP ® 180RTemperature Measurement SpecificationsProperty Description / ValueMeasurement range -40 ... +80 °C (-40 ... +176 °F)Accuracy over temperature range:at +15 ... +25 °C (+59 ... +77 °F) ±0.2 °C (±0.36 °F)at 0 …+15 °C and +25 …+40 °C(+32 ... +59 °F and +77 ...+104 °F) ±0.25 °C (±0.45 °F)at -40 ... 0 °C and +40 ... +80 °C(-40 ... +32 °F and +104...+176 °F) ±0.4 °C (±0.72 °F)Temperature sensor Pt1000 RTD 1/3 Class B IEC 751Operating Environment SpecificationsProperty Description / ValueOperating temperature rangeTransmitter body, no displayTransmitter body, with displayHMP110 probe-40 ... +60 °C (-40 ... +140 °F)-20 ... +60 °C (-4 ... +140 °F)-40 ... +80 °C (-40 ... +176 °F)Storage temperature range -50 ... +70 °C (-58 ... +158 °F)Electromagnetic compatibility EN 61326-1:2006 Electricalequipment for measurement, controland laboratory use - EMCrequirements - Basic immunity testrequirements.EN 55022:2006 + Am 1:2007Information technology equipment -Radio disturbance characteristics -Limits and methods ofmeasurement. Class B.