10GETTING STARTEDunpacking a new sensorWhen receiving a new microsensor remove the shock-absorbinggrey plastic net Please do not remove the seal and protectivetube before the following steps are successfully completed.connecT The TemperaTure sensorConnect the Unisense temperature microsensor to the inputterminal of the T301 or the Microsensor Multimeter or MonometercalibraTionWithout calibration, the Unisense temperature sensor incombination with the Unisense T301 thermometer has a precisionof ±1oC To verify the function and/or to achieve an even betterprecision (down to ±0 1 oC), perform the following calibrationprocedure: Prepare two solutions with well defined temperature,one slightly above and one slightly below the temperature rangein which the measurements are expected to be performedTo ensure that the calibration temperatures are well-defined,use for instance boiling water versus ice-water, or monitor thetemperature of the solutions with an independent thermometerwith a better precision than 1oCMicrosensor Multimeter/Monometer: Perform the calibrations inthe software Please consult the relevant software manualT301: If the display of the T301 deviates from the independentthermometer, you can correct subsequent measurements byadding the difference to the display valueWARNINGDo not removeseal and protectiveplastic tube beforethe following stepsare successfullycompleted.The Unisense T301 thermocouplemeter which is used for the TP-MR andTP200 temperature microsensors. |