No. 2D201-177EN*F82(2) AnalysisOpen the error log file. The following logs are displayed.Read the error log referring to the example below.(a) Data structureXL30 xxxx 000000000 mmddhhmm xxxxxxxxxx xx x x x x xxxxx xxxxxx 00<1> <2> <3> <4> <5> <6> <7> <8> <9> <10>Error No.Date of error occurrenceError code(R status Information)Command to be executedOperating modePreset tube voltagePreset tube currentPreset focal spot sizePreset exposure timeActual exposure timeSpare