Performance Tests413; AWG2021 Service Manual# . (' !' $' ( L+*D%#@ @ 8;L . 8;L /"< < ( @ 4,;,+ ! (-!! ! ( )*F%1. Use the test hookup and test equipment settings from the previous check.2. Check offset accuracy: Push the AWG2021 MANUAL TRIGGER button, and check that thedisplayed step is Step 5. Check that the DMM voltage reading is in the range from +2.470 V to+2.530 V. Push the AWG2021 MANUAL TRIGGER button, and check that thedisplayed step is Step 6. Check that the DMM voltage reading is in the range from –0.005 V to+0.005 V. Push the AWG2021 MANUAL TRIGGER button, and check that thedisplayed step is Step 7. Check that the DMM voltage reading is in the range from –2.530 V to–2.470 V.3. Check Option 02: If the AWG2021 has Option 02, repeat the Gain Accuracyand Offset Accuracy Checks for the AWG2021 channel 2 (CH2).4. End procedure: Disconnect the DMM. .