Page 62 SITRANS LR 300 – INSTRUCTION MANUAL 7ML19985CL02mmmmmParametersDisplay and Reading Parameters (P060 to P063)P060 Decimal Position (F = 2)Defines the maximum number of decimal places used on the LCD.In RUN mode, the decimal position is adjusted to prevent the number of digits fromexceeding the display capabilities. To keep the decimal place from shifting, reduce thenumber of decimal places to that shown at 100%.For example, if 100% = 15m, use two decimal places for readings of 15.00 or parts thereof(e.g. 12.25).P062 Offset Reading (F = 0.0000)Adds the specified value to the reading displayed, usually to reference the reading to sealevel or to another datum level.The operation of the device is not affected by the Offset Reading. This value is used fordisplay only. All control measurements are still referenced to empty.P063 Minimum Reading (F = 0.0000)Adjusts the minimum reading the product will show. This is useful on conical or parabolictanks to prevent negative values on the display.P063 is only applied after P062 Offset Reading and then only if reading displays level/volume (P001=1).The milliAmp output is not affected.Values (Levelmeasurement only)0 no digits after the decimal point1 1 digit after the decimal point2 2 digits after the decimal point3 3 digits after the decimal pointValues (Levelmeasurement only)Range: -999 to 99999Pre-set: 0.0000Values (Levelmeasurement only)Range: -999 to 99999Pre-set: 0.0000