33PBD-51041061 SITRANS LG260 - Operating Instructions44575-EN-200212Parameter settings for recording the echo curve and the settings ofthe echo curve itself can be carried out in the adjustment softwarePACTware.With the adjustment software PACTware and the PC the high-reso-lution echo curve can be displayed and used later on to assess thequality of the measurement.6.4.4 Additional adjustmentsIn this menu item, the internal clock of the sensor is set.After a reset, certain parameter adjustments made by the user arereset.Note:After this menu window, the reset process is carried out. No furthersafety inquiry follows.The following reset functions are available:Delivery status: Restores the parameter settings at the time of ship-ment from the factory, incl. order-specific settings. Any stored falsesignal suppression or user-programmed linearisation curve, as wellas the measured value memory, are deleted.Basic settings: Resetting of the parameter settings incl. specialparameters to the default values (presettings) of the respective instru-ment. Any created false signal suppression or user-programmable lin-earization curve as well as the measured value memory are deleted.The following table shows the default values of the instrument. De-pending on the instrument version or application, all menu items maynot be available or some may be differently assigned:Date/TimeReset