E382 E392E 061 01 02 02 Siemens-Elema AB 6 - 39KION / KION-i Troubleshootingx x 231 ... 00 E7 ... CON: DIAG_ HERC_TIMEOUT Fault: KION Monitor do not respond during PuC or Calibration.Description: The KION Monitor do not respond to commands within one second whenperforming a PuC or Calibration.What to do:1. Change KION Monitor and check if error disappears2. Check the function of the monitor docking station e.g by connecting anexternal docking station to P83 on PC 1753 in the Battery drawer.3. Check/replace the CPS/IDS board.x 232 ... 00 E8 ...Sub data:CON:DIAG_PUC_BIBD_INTERNAL_LEAKAGEFault: Leakage detected or one of the pressure transducers is not working properlyduring the Bellows position sensor calibration in the PuC.Description: Refer to sub data belowWhat to do: Refer to sub data below.... 00 02 DGBD_WAIT_FOR_EMPTY_BAG The Breathing bellows is emptied with a drive gas flow of 30 l/min.The Breathing bellows is not detected to be in lowest position after 30 seconds, i.e. thePEEP Pressure - Insp. pressure < 17.5 cmH2O.What to do:1. Check the seal on the bellows base plate.2. Check/replace EMV 7.3. Check/replace the Drive gas module.... 00 05 DGBD_WAIT_FOR_FULL_BAG The Breathing bellows is filled with a flow of 27 l/min from the AIR/N2O or O2 gasmodules.The Breathing bellows is not in top position after 30 seconds, i.e. the Exp. Pressure -PEEP Pressure < 15 cmH2O.What to do: Check/replace the Breathing bellows (leakage).... 00 0B DGBD_WAIT_FOR_FULL_BAG_FINAL_CHECKA final check that the calibrated Bellows position sensor is working by moving thebellows to top position again.The Breathing bellows is filled with a flow of 27 l/min from the AIR/N2O or O2 gasmodules.The Breathing bellows is not detected to be in top position after 30 seconds, i.e. theExp. Pressure - PEEP Pressure < 15 cmH2O.What to do: Check/replace the Breathing bellows (leakage).x 235 ... 00 EB ... CON:DIAG_TIMETEST_CURRENT_MAX_LOADDescription : Incorrect SW version.What to do: Check the CON-SW version in component log and report to EM-HSC.x 236 ... 00 EC ... CON: DIAG_TIMETEST_TOTAL_MAX_LOAD Description : Incorrect SW version.What to do: Check the CON-SW version in component log and report to EM-HSC.