20 System Delivery ProcedureACUSON S2000 US15-101.812.01.02.02 Siemens05.10 UltrasoundPage 20 of 401. Press the control panel Release button at the middle of the control panel handle.Fig. 12: Control Panel Movement2. Verify the control panel can be raised and lowered.- Raise the control panel to its highest position.- Lower the control panel to its lowest position3. Verify that the control panel can be swiveled from left to right.- Swivel the control panel to the left.- Swivel back to the center locked position.- Swivel the control panel to the right.- Swivel back to the center locked position.Verify Transducer Initialization 0Use the following process to verify the transducers and the transducer ports:1. Verify one of the transducers work in the left transducer port, see (Verify the Left Trans-ducer Port / p. 21) for instructions.2. Using the same transducer that was verified in the left port, verify the function of thecenter transducer port, see (Verify the Center Transducer Port / p. 24) for instructions.3. Using the same transducer that was verified in the center port, verify the function of theright transducer port (Verify the Right Transducer Port / p. 26).4. Verify one at a time, each transducer shipped with the system in any of the verifiedtransducer ports.5. Leave the last transducer being verified connected to the transducer port.