Pioneer N283 User Manual
Cleaning the main device1. Turn the unit off. Unplug power cord from the main device.2. Wipe the outside surface of the main device with a clean damp cloth every few days to keep itdust-free. DO NOT clean using abrasive cleaners.Danger: DO NOT submerge the main unit or the AC/DC adapter orDC car adapter in water. Doing so will result in devicedamage.MaintenanceWarning: The nebulizer kit is only designed for using, and should beused, with this specified compressor. Use with a differentkind of compressor may impair the performance.1. The disposable nebulizer kit must be replaced after every 30days of use. You can purchase more nebulizer kits from yourdevice supplier.2. On the bottom of the main device there is a filter holder andan inlet air filter, which ensures that the mist generated by thedevice contains no extraneous substances. The frequency ofchanging the filter depends on the using conditions of thedevice (e.g. in more or less dusty atmosphere). The filtershould be replaced every 6 months or sooner when a graycoloration is observed. Additional filters can be purchasedfrom your device supplier. filter holderfilterTo change the filter:1.) Remove the filter holder located on the bottom of the main device. Pull out the old filter fromthe holder and discard it.2.) Push a new filter into position, and reinstall the filter holder back to the main device.3. Authorized personnel maintain the rest of this device only.10 |