EN-6 SCOUTTM SERIES BALANCES - SPX3.4 Weighing ModeThis mode is the factory default setting.1. If needed, press and hold Mode until [wWeIGH] (Weigh) is displayed.2. If required, place an empty container on the pan and press Tare.3. Add sample to the pan or container. The display shows the weight of the sample.3.5 Counting ModeThis mode counts large numbers of items based on the weight of a reference count.1. Place an empty container on the pan and press Tare.2. Press and hold Mode until [Count] (Count) is displayed. [CLr.PwW] (Clear Average PieceWeight, APW) will then display.If no APW exists, the balance will display [Put.10], proceed to step 5.3. Press No to use the stored APW. Proceed to step 7.4. Press Yes to establish an APW.5. The balance will then display the stored sample size, i.e. [Put 10]. Press No or Back totoggle the choices (5, 10, 20, 50 or 100).6. Put the indicated number of pieces on the pan then press Yes to calculate the APW. Thedisplay shows the piece count.7. Add additional pieces until the desired count is reached.8. To clear the stored APW press and hold Mode until [Count] is displayed. Press Yes when[Clr.pwW] is displayed.Note: Press Function to view the current APW.3.6 Percent ModeThis mode measures the weight of a sample as a percentage of a reference weight.1. If required place an empty container on the pan and press Tare.2. Press and hold Mode until [Percnt] is displayed. [Clr.ref] (clear reference) will thendisplay. If no reference weight exists, the balance will display [Put.ref], proceed to step 5.3. Press No to use the stored reference weight and proceed to step 6.4. Press Yes to establish a new reference. Balance will now display [Put.ref].5. Add the desired reference material to the pan or container. Press Yes to store thereference weight. The display shows 100%.6. Replace the reference material with the sample material. The display shows thepercentage of the sample compared to reference weight.7. To clear the stored reference press and hold Mode until [Percnt] is displayed. Press Yeswhen [Clr.ref] is displayed.Note: Press Function to view the current reference weight.