RC04-PCR-LA Specifications 7-1Chapter 7 SpecificationsContains a list of the specifications.7.1 SpecificationsSetting functionsItem Setting range Resolution Setting accuracyPower line abnormality simulationT1 x1 deg 0 deg to 360 deg 1 deg For T2 = T4 = 0,1 degFor T2 ≠ 0 or T4 ≠ 0,1 msx1 ms 0.0 ms to 999.9 ms 0.1 ms For T2 = T4 = 0,±(1 x10-3 +0.1 ms)For T2 ≠ 0 or T4 ≠ 0±(1 x10-3 +1 ms)T2 x1 0 ms to 9 999 ms 1 ms ±(1 x10-3 +1 ms)x10 0.00 s to 99.99 sT3 x1 0.0 ms to 999.9 ms 0.1 ms For T2 = T4 = 0,±(1 x10-3 +0.1 ms)For T2 ≠ 0 or T4 ≠ 0,±(1 x10-3 +1 ms)x10 0 ms to 9 999msT4 x1 0 ms to 9 999 ms 1 ms ±(1 x10-3 +1 ms)x10 0.00 s to 99.99 sT5 x1 0 ms to 9 999 ms 1 ms 1 cyclex10 0.00 s to 99.99 s 10 msN x1 0 cycle to 9 999 cycles 1 cycle 1 cyclex10 0 cycle to 99 990 cycles 10 cyclesx100 0 cycle to 999 900 cycles 100 cyclesV (T3) Same as output voltage settingrange0.1 VRPT 0 to 9998 times or ∞ Once Once