Setup 2-port configuration using single moduleIn this section, the procedure to setup a 2-port configuration is shown. If you are using a non-Keysight chassis andcontroller, the basic procedure of the setup will be the same as the steps below.For tips and hints of setting up the M937X in non-Keysight chassis or controller, refer to appendix.Verify and install the softwareIf you know that your system has the latest software installed, skip this step and install the hardware.1. Power up the controller.If you’re using a remote controller and you have installed the interface cable, you must power up thechassis BEFORE you power up the PC. When you power down your system, shut down the PC BEFOREyou power down the chassis.2. Use windows key to show control panel, and select “Uninstall a program” to show list of programs installed inthe system.1. Check the revision of the Keysight IO Libraries Suite (IOLS). It should be revision 16.3.17914.4 ornewer.2. Check the revision of the M937X. It should be higher revision than the firmware on the web.www.keysight.com/find/pxivna3. Install software if the system does not have the latest software.1. Install the Keysight IO Libraries Suite. Using either the USB or the downloadable file atwww.keysight.com/find/iosuite. Follow the installer prompts to install the IO libraries. Thissoftware must be installed at first.2. Install the network analyzer software. Using either the USB memory or the downloadable file atwww.keysight.com/find/pxivna, launch the installer. Follow the installer prompts to install soft frontpanel, device drivers (IVI-C, IVI-COM, and LabVIEW G) and documents.When you install the product software, the IVI driver files are installed in the standard IVI Foundationdirectories (for example, C:\Program Files\IVI Foundation\IVI\Drivers\). Example programs areprovided to demonstrate most driver functionality. The reference material for the driver functions (aMicrosoft HTML Help .chm file) is installed. In addition, you can directly access the .chm file(AgM9391.chm) from this Start menu location: Start > All Programs > Agilent IVI Drivers > AgNA >AgNA Documentation.5