Quick Demo GuideKeysight Technologies E4980A Precision LCR Meter, 20 Hz to 2 MHzE4980A Precision LCR MeterProcedure overview1. Preset system2. Perform compensation of fixture3. Set measurement parameters4. Connect DUT and measure at a single frequency5. Set up list sweep table6. Set trigger mode and make sweep measurementIn this demo...– Measurement of chip SMD device– Fixture compensation– List sweepRequired Instrument and fixtureFixture for SMD device(16034E/G/H, 16334A)If the fixture for SMD device is not available, it is pos-sible to perform the same procedure with followingfixtures. Use lead components in this case.16047D 16089A/B/C/D/E16047E1. Preset systema. Press [Preset] front panel keyb. Press CLEAR SET&CORR softkeyc. Press OK softkey (wait about 20 seconds…)2. Perform compensation of fixturea. Set fixture to UNKNOWN terminalsb. Press [Meas Setup] front panel keyc. Select CORRECTION softkeyd. Move cursor onto OPEN field by pressing[▼] front panel keye. Make open condition for the test fixtureTo make open condition, adjust the distance of thetwo electrodes so that the separation becomes equalto the length of the DUT (Figure 1)f. Press MEAS OPEN softkey to perform opencorrectiong. Press ON softkeyTo activate Open correction, the status of the OPENfield should be changed from “OFF” to “ON.” (Figure 2)Figure 1. Make open condition (for 16034E/G/H).Figure 2. Turned on OPEN correction.h. Make short condition for the test fixturei. Move cursor onto SHORT field by pressing[▼] front panel keyj. Press MEAS SHORT softkey to perform shortcorrectionk. Press ON softkeyTo make short condition, short the test fixture bysliding the right (HIGH) electrode to the left so thattwo electrodes contact to each other