Section 3: Adjustment Model DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual3-24 DMM6500-905-01 Rev. A/July 2018Figure 30: Connection for a 10 A range adjustmentCommand parameters for an AC current 10 A range adjustmentWhen calibrating your DMM6500 for an AC current 10 A range adjustment, use the followingcommand parameters.Send each command parameter twice. First, send the parameter using the setup command.Second, send the parameter using the execute command.The example code below uses the first line of code from the table below. Send all commandparameters using this command.cal.adjust.step.setup("cal_ACI_10A_400Hz_tenth")cal.adjust.step.execute("cal_ACI_10A_400Hz_tenth")Command parameters CalibratorfunctionCalibratorrangeCalibratorvalueCalibratorfrequencycal_ACI_10A_400Hz_tenth AC current 2.1 1 400 Hzcal_ACI_10A_400Hz_fs AC current 11 7 400 Hzcal_ACI_10A_5kHz_fs AC current 11 7 5 kHzcal_ACI_10A_400Hz_2A AC current 2.2 2 400 Hzcal_ACI_10A_10Hz_2A AC current 2.2 2 10 HzFrequency adjustmentThe following procedure provides instructions for completing a front-terminal adjustment forfrequency. The following section provides a command parameter table to complete the adjustment.