16 PARAMETER SETUP MENU ABSOLUTE EC RESOLUTION UNIT (HI763093 EC PROBE ONLY)Option: Auto, Auto mS/cm, 1μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm, 1 mS/cmPress the Modify key to change the value. Use the / arrow key to select the value. Press Selectto confirm or ESC to return to the menu without saving.Absolute conductivity displays the conductivity measurement without temperature compensation. Asmall letter “A” is added to the measurement unit to indicate no temperature compensation.TDS RESOLUTION UNIT (HI763093 EC PROBE ONLY)Option: Auto, Auto ppt, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt, 1 pptPress the Modify key to change the TDS resolution. Use the / arrow key to select the resolution.Press Select to confirm or ESC to return to the menu without saving.Auto allows the meter to choose the range that will optimize the measurement.Auto ppt allows the meter to choose the range that will optimize the measurement, reading will onlybe in ppt.1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt or 1 ppt the meter will display the measurement in the selectedresolution.DO CONCENTRATION UNIT (HI764103 DO PROBE ONLY)Option: ppm or mg/LPress the functional key to select the desired dissolved oxygen measurement unit.7.3. PARAMETER COEFFICIENTS (EC & DO probes only)From the Parameter setup menu, use the / keys to highlight Parameter coefficients and thenpress Select.Warning: The previously saved logs for the connected probe will be updated with the selectedcoefficients.