1918ACCESSORIESK-TYPE THERMOCOUPLE PROBESwith integral handle, 1 m (3.3') cable & mini-connectorHI 766A Roller surface probe, max 320°C/600°FHI 766B Surface probe, max 650°C/1200°FHI 766B1 90º surface probe, max 450°C/840°FHI 766B2 Spring-loaded, surface probe, max 900°C/1650°FHI 766B3 Spring-loaded, small surface probe w/insulated shaftHI 766C Penetration probe, max 900°C/1650°FHI 766C1 Ultra-fast Penetration probe, max 300°C/570°FHI 766D Air probe, max 300°C/570°FHI 766E1 General purpose probe, max 900°C/1650°FHI 766F High temperature, flexible wire probe without handleHI 766F1 Flexible wire probe w/o handle, max 480°C/900°FHI 766TR1 Penetration probe, max 250°C/482°FHI 766TR2 Penetration long probe, max 250°C/482°FHI 766TV1 Pipe clamp probe, max 200°C/390°Fwith detachable handle & mini-connector (to be used in con-junction with the HI 766HD probe handle)HI 766PA Roller surface probe, max 320°C/600°FHI 766PB Surface probe, max 650°C/1200°FHI 766PC Penetration probe, max 900°C/1650°FHI 766PD Air probe, max 300°C/570°FHI 766PE1 General purpose probe, max 900°C/1650°Fgrill surface probeHI 766B4 Grill surface probe with 70 cm (27.6’’) cable (protectedwith stainless steel jacket), max 250°C/482°FHI 7664B4S Spare stainless steel sensor for HI 766B4 probeOTHER ACCESSORIESHI 710007 / 8 Shockproof rubber boot, blue / orangeHI 710018 Spare protective caseHI 766EX Extension cable for K-type probesHI 92000 Windows® compatible softwareHI 920011 Serial cable for PC connectionRecommendations for UsersBefore using these products, make sure that they are entirely suitable for theenvironment in which they are used. Operation of these instruments in residentialarea could cause unacceptable interferences to radio and TV equipments, requiringthe operator to take all necessary steps to correct interferences. Any variation intro-duced by the user to the supplied equipment may degrade the instruments' EMCperformance. To avoid electrical shock, do not use these instruments when voltagesat the measurement surface exceed 24 Vac or 60 Vdc. To avoid damages or burns,do not perform any measurement in microwave ovens.USER CALIBRATION(Not for HI 93530, HI 93531 and HI 93532 models)The meters can be calibrated at 0°C by using an ice bath.• Prepare an ice bath with approximately equal volumes of distilledwater and chopped ice made from distilled water.• Immerse the temperature probe in the center of the ice bath,taking care not to touch the ice with the probe tip.• Ensure that the meter is measuring a temperature within ±3°C.• To enter the Calibration mode, press and hold the CAL button forabout 5 seconds.• The CAL tag turns on to indicate that theCalibration mode has been entered.Note: If the measurement is outside the±3°C window, the meter does notenter Calibration mode.• When the meter reaches the stability condition, which is detectedwhen the measurement remains within ±0.2°C for 5 seconds,the calibration is accepted and the reading becomes 0°C (32°F).• The meter then automatically returns to normal mode.Note: To exit the Calibration mode at any time, press the CAL key.Note: User calibration cannot be entered in T1-T2 mode.Note: User calibration is only performed on the current channeldisplayed (T1 or T2).FACTORY RECALIBRATIONAll Hanna thermometers have been accurately pre-calibrated at thefactory.It is generally recommended to have all thermometers recalibrated atleast once a year.For an accurate recalibration, contact your nearest Hanna CustomerService Center.Note: To clean the meters, do not use aggressive detergents. It isrecommended to use water.