PARAGRAPHNUMBER TEST TEST LIMIT TESTRESULT5-12 AC Voltage Accuracy200 mV Range40 Hz to 10 kHz10 mV 933 - 10.07 mV50 mV 49.87 --50.13 mV100 mV 99.80 --100.20 mV10 kHz - 20 kHz10 mV 9.80 - 10.20 mV50 mV 49.60 - 50.40 mV100 mV 99.35 - 100.65 mV2 V Range40 Hz - 10 kHz0.1 V .0993- .1007 V0.5 V .4987 - .5013 V1.0 V .9980 - 1.0020 V10 kHz - 20 kHz0.1 V .0980 - .1020 V0.5 V .4960 - .5040 V1.0 V .9935 - 1.0065 V20 V Range40 Hz - 10 kHz1 V .993 - 1.007 V5 V 4.987 --5.013 V10 V 9.980 -- 10.020 V10 kHz - 20 kHz1 V .980 - 1.020 V5 V 4.960 - 5.040 V10 V 9.935 - 10.065 V200 V Range40 Hz - 10 kHz10 V 9.93 - 10.07 V50 V 49.87 -50.13 V100 V 99.80 - 100.20 V10 kHz - 20 kHz10 V 9.80 - 10.20 V50 V 49.60 - 50.40 V100 V 99.35 - 100.65 V500 V Range40 Hz - 1 kHz100 V 99.3 - 100.7 V500 V 498.7 - 501.3 V1 kHz -2 kHz100 V 99.0 - 101.0 V500 V 497.0-- 503.0 V