Data Management Center (DMC)LOGIQ 500 Advanced Reference Manual2276614–100 Rev. 04–32 Error MessagesIf an anomaly occurs in the communication or transfer of databetween the LOGIQ 500 and the PC, an error message willbe displayed on the LOGIQ 500 screen. Possible errormessages are:Message Possible CauseCheck Computer. No power. PC power is off. Interface cable broken or discon-nected.Check Computer. Bad condition. PC is busy. LOGIQ 500 senses the PC cannotaccept data.Check Computer. Abnormal response. LOGIQ 500 senses an abnormal response fromthe PC.Check Computer. Checksum Error occurred. LOGIQ 500 senses an abnormal acknowledge-ment from the PC after all data has been sent.Check Computer. No response. LOGIQ 500 senses noacknowledgement from the computer after all datahas been sent.Can’t send data to Computer in this category. User is trying to send data to the PC in an examcategory other than OB, GYN or Vascular.System doesn’t have option. The OB calculation option for the LOGIQ 500 hasbeen turned off.SYSTEM ERROR: CANNOT COMMUNICATEWITH COMPUTER.A LOGIQ 500 system error has occurred that willnot allow communication with the computer to begin.This function is not available. This function is not available in some situations.Table 4–1. Data Management Center Error Messages